[Objective]
Objective of electron cyclotron
emission (ECE) diagnostic systems is to obtain time evolution of electron temperature
with high temporal (~microsec) and spatial (~1 cm) resolutions.
The ECE diagnostic systems are
composed of the following three systems:
Fourier Transform Spectrometer system (FTS),FTS is absolutely calibrated using liquid nitrogen, and GPS and HRS are relatively calibrated using the FTS data.
Grating Polychromator System (GPS),
Heterodyne Radiometer System (HRS).
[Specifications]
Antennas for the ECE measurement
are located at P14 and P17 sections, which are 60 degrees apart in toroidal direction.
Transmission line in the torus hall is shown in Fig. 1.
ECE from the JT-60U plasma is transmitted
through corrugated circular waveguides (diameter: 63.5 mm) and smooth-wall rectangular
waveguides (WR-284). The transmission line, whose total length is about 40 m,
goes to the diagnostic room outside the torus hall.
Transmission line in the diagnostic room is shown in Fig. 2. Two power dividers are used to lead the ECE to FTS, GPS and HRS. A highpass filter to reject electron cyclotron heating (ECH) wave (f=110 GHz) is also inserted. In addition, grating lowpass filters are installed in the GPS transmission line to reject spurious ECE.
ECE ranging from 100 to 300 GHz is measured with these systems, and second harmonic ECE is used for electron temperature measurement. Specifications of the ECE systems are shown in Table 1.
Fig. 1: ECE
transmission line in the torus hall.
Fig. 2 : | Layout of ECE diagnostic systems in
the diagnostic room. Transmission line from the P17 port is connected to the HRS transmission line depending on experimental schedule. |
Table 1: Specifications of FTS, GPS and HRS
|
|
|
|
Temporal resolution (max.) |
|
|
|
Frequency range [GHz] |
|
|
102-124 |
Frequency resolution [GHz] |
|
|
|
Spatial resolution (for B=4T) [cm] |
|
|
|
Number of channels |
|
|
|
Calibration |
|
|
|
[ECE data analysis]
Analysis using ECE data has been
extensively performed in collaboration with Daido
Institute of Technology. Research topics are as follows:
- Effect of relativistic frequency down-shift on ECE measurement
- Adaptive neural network for detection of a sawtooth crash
- Singular value decomposition for separation of perturbative temperature component
- Spectrum estimate from FTS interferograms using maximum entropy method
References
* Review of ECE Diagnostic
Systems
[1] M. Sato, S. Ishida,
N. Isei, A. Isayama, H. Shirai, T. Oyevaar, M. Teranishi, N. Iwama
and K. Uchino:
'Measurements and Analysis
of Electron Cyclotron Emission in JT-60U', Fusion Eng. Design 34-35
(1997) 477.
[2] N. Isei, A. Isayama, S. Ishida, M. Sato,
T. Oikawa, T. Fukuda, A. Nagashima, N. Iwama and JT-60 Team:
'Electron Cyclotron
Emission Measurements in JT-60U', Fusion Eng. Design 53
(2001) 213.
* Fourier Transform Spectrometer
System
[3] M. Sato, H. Yokomizo
and A. Nagashima:
'Fourier Transform Spectrometer
System on JT-60', Kakuyugo Kenkyu (J. Plasma Fusion Research) Supplement, 59 (1988) 47 [in Japanese].
* Grating Polychromator System
[4] S. Ishida, A. Nagashima, M. Sato, N. Isei and T. Matoba:
'Twenty-Channel Grating Polychromator
Diagnostic System for Electron Cyclotron Emission Measurements', Rev.
Sci. Instrum. 61 (1990) 2834.
[5] A. Isayama, N. Isei,
S. Ishida and M. Sato:
' A 20-channel Electron
Cyclotron Emission Detection System for a Grating Polychromator
in JT-60U', Rev. Sci. Instrum. 73 (2002) 1165.
* Heterodyne Radiometer System
[6] N. Isei, M. Sato, S.
Ishida, K. Uchino, A. Nagashima, T. Matoba and T. Oyevaar:
'Development of 180 GHz Heterodyne Radiometer for Electron Cyclotron Emission Measurement in JT-60U', Rev. Sci. Instrum. 66 (1995) 413.
[7] A. Isayama, N. Isei,
S. Ishida, M. Sato, Y. Kamada, S. Ide, Y. Ikeda, K. Takahashi, K. Kajiwara, K. Hamamatsu and JT-60
Team:
'Electron Temperature
Perturbations Measured by Electron Cyclotron Emission Diagnostic Systems in JT-60U',
Fusion Eng. Design 53 (2001) 129.
* Grating filter for GPS
[8]
M. Sato, N. Isei and S. Ishida:
'Grating Filter for
Grating Polychromator on Measurement of Electron Temperature Profile
from Second Harmonic Electron Cyclotron Emission', J. Plasma Fusion Res. 71 (1995) 748
[in Japanese].
* ECE Data Analysis
[9] M. Sato, S. Ishida and
N. Isei:
'Relativistic Broadening
Effect on Application of Electron Cyclotron Emission Measurements
to High Temperature Tenuous Tokamak Plasma', J. Phys. Soc. Japan 62 (1993) 3106.
[10] M. Sato, N. Isei and
S. Ishida:
'Relativistic Down-Shift
Frequency Effect on the Application of Electron Cyclotron Emission Measurements to Medium
Temperature Tokamak Plasmas', Jpn. J. Appl. Phys. 34 (1995) L708.
[11] M. Sato, N. Isei, S.
Ishida and A. Isayama:
'Effects of Relativistic
Frequency Down-Shift and Optical Thickness on Measurements of Electron Temperature Profile
from Electron Cyclotron Emission in Medium Temperature Tokamak
Plasmas', J. Phys. Soc. Jpn. 67 (1998) 3090.
[12] A. Isayama, Y. Kamada,
T. Ozeki and N. Isei:
'Measurement of Magnetic
Island Width in Long-pulse, High-beta_N Discharges in JT-60U', Plasma Phys. Control.
Fusion 41 (1999) 35.
[13] M. Sato, N. Isei, A.
Isayama and S. Ishida:
'Effects of Polarization
Change on the Measurement of Electron Temperature Profile from
ECE in Tokamak Plasma', Fusion Eng. Design 53 (2001) 161.
[14] A. Isayama, N. Iwama,
Y. Hosoda, S. Satake, N. Isei, S. Ishida and M. Sato:
'Singular Value Decomposition
Analysis of Multichannel Electron Cyclotron Emission Signals of Tokamak Plasma',
Jpn. J. Appl. Phys. 42 (2003) L329.
[15] A. Isayama, N. Iwama,
T. Showa, Y. Hosoda, N. Isei, S. Ishida and M. Sato:
'Maximum Entropy Estimation
of Electron Cyclotron Emission Spectra from Incomplete Interferograms in ELMy
H-mode Tokamak Experiment', Jpn. J. Appl. Phys. 42 (2003) 5787.
[16] M. Sato, A. Isayama, N. Iwama and K. Kawahata:
'Feasibility of Electron Density Measurement Using Relativistic Downshift of Electron Cyclotron Emission in Tokamak Plasmas', Jpn. J. Appl. Phys. 44 (2005) L672.
[17] A. Isayama, N. Oyama, H. Urano and the JT-60 Team:
'Measurement of Toroidal Structure of Electron
Temperature with Electron Cyclotron Emission Diagnostic in JT-60U', Proc. 21st IEEE/NPSS Symposium
on Fusion Engineering 2005 (SOFE2005).