Vertical analyzer |
Tangential analyzer | ||
-Maximum energy | : | 4 MeV for ![]() 2 MeV for H |
4 MeV for ![]() 4 MeV for H |
-Energy ratio | : | 4 | 3.85 |
-Number of detectors | : | 8 | 8ch ![]() |
-Scintillator thickness | : | 10 mm | 2-23 mm |
-Energy resolution | : | 6-11% | 6-11% |
-Position | : | P-12 section | P-8 section |
![]() |
Tangential analyzer |
![]() |
Vertical analyzer |
Development, installation and commissioning under the collaboration between JAERI and the Ioffe Institute, Russia.
[Reference]
(1) Y. Kusama et al., "Charge-exchange neutral particle measurement
in MeV energy range on JT-60U", Rev. Sci. Instrum. 66(1)
339(1995).