Visible spectrometer for divertor


[Objectives]
To observe spectral lines, bands and their spatial profile for study of atomic and molecular processes in the divertor plasmas.

[Diagnostic method]
Emission collimated with lenses is transmitted through 16 optical fibers from the torus hall to a diagnostic room. The light from the optical fibers is dispersed with the spectrometer and the image-intensified CCD camera observes the image of the fibers.

[Specifications]
Spectrometer 1
Spatial resolution : 2 - 4 cm ( 16 ch )
Focal length : 0.5 m ( F/6.5 )
Gratings : 150 g/mm, 600 g/mm and 2400 g/mm
Wavelength resolution : 1.3 nm, 0.29 nm and 0.054 nm
Wavelength band : 150 nm, 37 nm and 7 nm
Time resolution : 25 ms
Position : P-8 section upper

[Reference]
T.Nakano et al., JAERI-Research 2000-35 (2001).